We established three optimal cutoff points to predict Direct Gram stain positive samples for peritoneal (465.0 bacteria/μL), synovial (1200.0 bacteria/μL), and cerebrospinal fluids (17.2 bacteria/μL) with maximum sensitivity and negative predictive values. ![]() Moreover, the bacterial count obtained with UF4000 (Sysmex, Kobe, Japan) was significantly higher among culture or Direct Gram stain positive samples. With regard to cell counts, UF4000 (Sysmex, Kobe, Japan) showed a performance that was at least equivalent to those of the reference methods and superior to those of XN10 (Sysmex, Kobe, Japan). All samples were analyzed by flow cytometry and subsequently processed following cytological and/or microbiological routine procedures. A total of 526 biological fluid samples were included in this study (42 ascitic, 31 pleural, 31 peritoneal, 125 cerebrospinal, 281 synovial, and 16 peritoneal dialysis fluids). We evaluate the analytical performance of SYSMEX UF4000 (Sysmex, Kobe, Japan) and Sysmex XN10 (Sysmex, Kobe, Japan) in comparison with cytological and microbiological routine procedures. The microbiological routine procedures consist of Direct Gram staining and culture on solid or liquid media. ![]() The gold standard method for cell count in biological fluids is the manual method using counting chambers. The analysis of biological fluids is crucial for the diagnosis and monitoring of diseases causing effusions and helps in the diagnosis of infectious diseases.